Hach DR 1900 Portable Spectrophotometer
Carry more water methods with you wherever you need them most.
The DR1900 excels in the field because it is the lightest and most compact portable spectrophotometer. Your field testing takes you to potentially dusty and wet conditions where other equipment simply won’t be safe. The DR1900 is built for rugged conditions, and it’s flexible, –accepting the widest range of vial sizes. Built with field use in mind, the DR1900 has a large, clear screen and a simple user interface that makes testing easier than ever in even the most demanding conditions. Easy to hold, and operate, the compact DR1900 is a valuable tool for your field technicians.
Underneath the rugged exterior, the DR1900 has the largest number,– over 220– of the most commonly tested preprogrammed methods already built in. You can also use the easy-to-use interface to create your own methods. Tests are performed with a wavelength range of 340 to 800 nm which make this a field instrument you can use to find results typically only seen in laboratory instruments.
- Carry Anywhere
- Test What You Need
- Ensure Accuracy in the Field
- Make Testing Easier
- Take Into Field Environments
Specifications:
- Battery Life:15 days (typical) @ 5 readings a day / 5 day week without backlight*
* Backlight usage will decrease battery life. - Battery Requirements:4, AA size alkali cells
- Cuvette Compatibility 2:10 mm square / 1 inch square and 13 mm / 16 mm / 1 inch round
- Data Logger:500 measured values (Result, Date, Time, Sample ID, User ID acc. to GLP)
- Detector:Silicon photodiode
- Dimensions (H x W x D):98 mm x 178 mm x 267 mm
- Display:Graphical display 240 x 160 pixel (LCD, b/w, backlit)
- Enclosure Rating:IP67
- Includes:Printed Basic Manual, CD with procedure manual (English) in pdf format. Set of vial adapter in carton box. Pkg of 4 AA Alkaline Battery. Dust Cover
- Interface:USB type Mini IP67 (with optional Module)
- Mains Connection 2:110 – 240 V / 50/60 Hz (* optional Module required. May not be available in all regions)
- Mains Connection 3:4x NiMH rechargeable Battery (* optional Module required. May not be avaliable in all region)
- Operating Conditions:10 to 40 °C (50 – 104 °F), max. 80 % relative humidity (non-condensing)
- Operating Humidity:max. 80 % relative humidity (non-condensing)
- Operating Mode:Transmittance (%), Absorbance and Concentration
- Operating Temperature:10 to 40 °C
- Optical System:Reference beam, spectral Reference beam, spectral
- Photometric Accuracy:± 0.003 Abs @ 0.0 – 0.5 Abs
- Photometric Linearity:< 0.5 % (0.5 – 2.0 Abs)
- Photometric Measuring Range:0 to 3 Abs (wavelength range 340 – 800 nm)
- Power supply:4 x AA size Alkaline
4x NiMH rechargeable Battery*
Power supply*: 110 – 240 V; 50/60 Hz
(* optional Module required. May not be available in all regions) - Preprogrammed Methods:> 220 (Please note: TNT vial test evaluation possible, but without barcode reading & 10 fold measurement and outlier elimination feature.)
- Reproducibility:± 0.005 Abs (0 – 1 A)
- Sample Cell Compatibility:13 / 16 mm and 1 inch round adapter, 10 x 10 mm 1 inch square and 10x10mm
- Source Lamp:Xenon Flash
- Spectral Bandwidth:5 nm
- Storage conditions:-30 to 60 °C (-22 to 140 °F), max. 80 % relative humidity (non-condensing)
- Stray Light:< 0.1 %T at 340 nm with NaNO2
- Temperature Range:0 to 50 °C (32 – 122 °F)
- User Interface Languages:English, French, German, Italian, Spanish, Portuguese, Bulgarian, Chinese, Czech,
Danish, Dutch, Finnish, Greek, Hungarian, Japanese, Korean, Polish, Romanian,
Russian, Slovenian, Swedish, Turkish - User Programs:50
- Wavelength Accuracy:± 2 nm (range 340 – 800nm)
- Wavelength Calibration:Automatic
- Wavelength Range:340 to 800 nm
- Wavelength Reproducibility:± 0.1 nm
- Wavelength Selection:Automatic
- Weight:1.5 kg